Richards Thermocouple and Temperature Sensor Catalog 16
Manufacturing, Engineering and Sales Since 1938
ASR900
Phone: 617.527.4385 Fax: 617.964.3746 Email: sales@asrichards.com
Digital Temperature Controller - ASR900HY
The Richards model ASR900HY micro processor temperature process controller is a very economical solution for temperature control in your process. These instruments are also inexpensive enough that they can also be used in situations where just an indicator is required. They are accurate and have selectable inputs for all standard thermocouple Types (B, C, E, J, K, N, R, S, T, & Platinel II) and Pt100Ώ & KPt100Ώ three wire RTDs. Their sizes are excellent for panel mounting of multiple instruments. The ASR900HY is a bezel mount instrument, which require no mounting screws for installation in panels.
√ Fuzzy Function, PID auto tuning
√ Control Outputs: Relay, SSR, & SCR
√ RS485/422 communication function
Features:
Example range: Type K (Thermocouple): -200 to 1370°C (0 to 2498°F )
ASR900HY - 1/4 DIN Panel Mount
1
Item
General Specifications
Power supply
100-240 VAC, 50-60 Hz
Power Consumption
10 VA max.
±0.5% of FS ±1 Digit, thermocouple (B, E, J, K, L, R, S, T, U, W, PL2) ±01.0% of FS ±1 Digit, thermocouple (N)
Display Accuracy
±0.5% of FS ±1 Digit, RTD (KPt100Ώ, Pt100Ώ)
Sampling Time
250 ms
Relay: Contact capacity: 1C, 240V AC, 3 A, 30V DC, 3A (resistive load)
SSR: More than approx. 12V DC (Min 600Ώ resistive load), with disconnection, limit within approx. 30 mA, Time resolving power: smaller one between 0.1% and 10 ms SCR: 4 - 20 mA DC (resistive load less than 600Ώ) Accuracy: ±0.5% of FS (range 4 - 20 mA DC) resolving power : approx. 3,000 2,300 V AC, 50 / 60 Hz, for 1 min (1st terminal - 2nd terminal - between the earth terminal) 1,500 V AC, 50 / 60 Hz, for 1 min (2nd terminal - between the F.G)
Control Output
Instrument Rear Panel - Hookups
1
11
OUT1
2
12
Dielectric strength
3
13
OUT1 SSR / SCR RET
4
14
Min 20 MΏ (500V DC), 1st terminal - 2nd terminal - between the earth terminal
Insulation Resistance
5
15
OUT2 SSR / SCR RET
6
16
Ambient temperature
0 to 50°C
RTD
Storage Temperature
-25 ~ 65°C
7
17
A
Ambient Humidity Range
8
18
35 ~85 % RH (without dew condensation)
~
Power
B
9
19
Net Weight
Approximately 472 grams
/ VVVV\
10
20
Panel Cut-out
92.0 mm x 92.0 mm (3.62” x 3.62”)
mV / V
B
Instrument Outline
96 mm x 96 mm x 112.5 mm (3.78” x 3.78” x 4.43”)
Thermocouple
Allowable wiring resistance
RTD (max 10Ώ, however resistance among 3 wires should be same)
18
TC
ASR900HY
Cold junction compensation error
±1.5°C (15 ~ 35°C), ±2.0°C (0 ~ 50°C)
19
20
Input signal break detection
OFF, UP/DOWN Scale selection (Thermocouple), UP Scale (RTD)
Red
Visit our website at www.asrichards.com
Page 175
Made with FlippingBook